X-ray fluorescence (XRF) spectrometers SuperMini 200
The new Supermini200 has improved software capabilities as well as a better footprint. As the world's only high-power benchtop sequential wavelength dispersive X-ray fluorescence (WDXRF) spectrometer for elemental analysis of oxygen (O) through uranium (U) of almost any material, the Rigaku Supermini200 uniquely delivers low cost-of-ownership (COA) with high resolution and lower limits-of-detection (LLD).
  1. Features
  2. Specification

Low cost of ownership 

Not only is the Rigaku Supermini200 an affordable choice for your XRF elemental analysis needs, the running costs are also low. P10 gas for the flow proportional detector is the only consumable. The Supermini200 does not require a source of cooling water, plumbing, or an external chiller, thereby decreasing system maintenance and lowering the overall lifetime cost of ownership of the system and yearly budgets for consumables and maintenance.

Superior light element performance

Analyzing complex matrix materials with a wide range of light and heavy elements, from trace to high concentration levels, is the instrument's core competency. Rigaku Supermini200 delivers high sensitivity for light elements with superior spectral resolution for resolving line overlaps in complex matrices without the need for deconvolution. Analyzing low concentration levels of light elements (F, Na, Mg, Ca, Si, Al, and P) is easy.

High power X-ray tube 

Supermini is the only high powered (200 W) X-ray tube benchtop WDXRF system, providing excellent excitation resulting in lower detection limits and shorter measurement times. A unique and proprietary optical system is also employed to enhance sensitivity beyond the gains of the tube power.

Features

  • Analyze oxygen through uranium (O → U)
  • Analyze: solids, liquids, powders, alloys and thin films
  • Atmosphere: air, helium or vacuum
  • X-ray tube: 50 kV, 200 W Pd-anode
  • Primary beam filter: Zr is standard; Al optional
  • Detectors: F-PC and scintillation
  • Crystals: 3-position changer
  • Autosampler: 12-position standard
  • Vacuum: rotary pump standard
  • Power: 100 – 120V (50/60 Hz) 15A or 200 – 240V (50/60 Hz) 10A

X-ray generator

Max. 50 kV, 4 mA

X-ray tube

Pd target, 200 W air-cooled

Crystal changer

3 crystal changer

Analyzing crystals

Standard: LiF 200 (for Ti-U) and PET (for Al-Ti), option: RX25 (for F-Mg) or Ge (for P-Ca)

Detectors

Light elements: F-PC (gas flow proportional counter), heavy elements: SC (scintillation counter)

Primary beam filter

Automatic in and out; select one from Zr filter for Cd analysis (standard) and Al filter for trace K analysis

Sample changer

12-position sample turret

Sample size

Maximum sample size 44 mm diameter, 33 mm height

Sample spinning

30 rpm

Spectrometer chamber atmosphere

Vacuum, vacuum/helium (option)

Vacuum pump

Rotary pump

Operating system

Windows® 7

Power requirements

100 – 120V (50/60 Hz) 15A or 200 – 240V (50/60 Hz) 10A

Installation environment

Temperature 15-28°C, relative humidity less than 75%