The Extraction System B-811 is the most convenient and versatile of all automated extraction systems offering 4 different extraction techniques in one unit. Perform a Soxhlet Standard, a Soxhlet Warm, a Hot Extraction, or Continuous Flow – each process is fully automated and controlled. Additionally: work under completely inert conditions!The instrument is available in a standard and LSV configuration for large volume samples (trace analysis)
Hot Extraction, Soxhlet Standard, Soxhlet Warm and Continuous Flow can be carried out with the B-811
Automatic sequencing: The electronic unit controls the sequence of the three phases (extraction, rinsing, drying) of between one to four samples in parallel.
Quicker execution of the extraction process: Two built-in heating sources per extraction location and optimal heat transfer permit the use of high-boiling point solvents. Moreover, control by means of the optical sensor makes it possible for the user to optimize the exchange of the solvent from a timing point of view, to increase the material exchange (analyte/solvent) and, at the same time, to economize in solvent consumption.
Better operating protection thanks to higher safety standards: Cooling water, level of solvent and heat source are directly monitored, the complete apparatus is specially produced to view complete tightness of joints while a protective shield provides an extra precaution.
Sample-compatible drying: In the case of heat-sensitive samples, the extract can be concentrated in a protective mode and oxidation-sensitive samples can additionally be extracted and dried in an inert gas atmosphere.
Sample purity is maintained at all times: All components in contact with the sample are inert so the possibility of contamination of the sample
Mains Voltage: 230V
Frequency: 50/60 Hz
Capacity: 1250 W
Volume of extraction vessel: 120 ml (220 ml for B-811 LSV)
Length of condenser tube: 312 ml (291 ml for B-811 LSV)
Diameter of sample holder: 39 ml (49 ml for B-811 LSV)
Volume of glass sample thimble: 150 ml (280 ml for B-811 LSV)
Volume of solvent beaker: 150 ml (230 ml for B-811 LSV)