Features |
Polarized Cartesian geometry for trace level sensitivity
Unlike conventional EDXRF analyzers, the NEX CG was engineered with a unique close-coupled Cartesian Geometry (CG) optical kernel that dramatically increases signal-to-noise. By using secondary target excitation, instead of convention direct excitation, sensitivity is further improved. The resulting dramatic reduction in background noise, and simultaneous increase in element peaks, result in a spectrometer capable of routine trace element analysis even in difficult sample types.
Novel software reduces the need for standards
NEX CG is powered by a new qualitative and quantitative analytical software, RPF-SQX, that features Rigaku Profile Fitting (RPF) technology. The software allows semi-quantitative analysis of almost all sample types without standards — and rigorous quantitative analysis with standards.
Features
- Analyze 11Na to 92U non-destructively
- Solids, liquids, powders and thin films
- Polarized excitation for lower detection limits
- Novel treatment of peak overlap reduces errors
- PPB detection limits for aqueous samples using UltraCarry
- Simplified user interface with EZ Analysis
|
Specification |
Excitation
- X-ray tube with Pd anode
- 50W max power
- 50kV max voltage
- Four standard polarization and secondary targets depending
on application, for optimum excitation
- Optional fifth target for optimal excitation of Na and Mg
Detector
- High performance SDD
- Peltier electronic cooling
- Large active detection area
- Optimum balance of spectral resolution and high
count rate
Sample Chamber
- Large 38 cm diameter x 10 cm deep sample chamber for bulk
samples
- 15-position automatic sample changer (32mm sample cups)
- 10-position automatic sample changer (35-40mm sample cups
and pellets)
- 9-position automatic sample changer with sample spinner
- Analysis in air, helium purge, or vacuum
Environmental Conditions
- Ambient temperature 18-28ºC (65-82ºF)
- Relative humidity ≤75%
- Vibration: undetectable by people
- Free from corrosive gas, dust and particles
|
Computer
- External PC computer system
- Microsoft® Windows Vista® operating system
- Keyboard and mouse
- Monitor
- Printer
Software & Application Packages
- Menu-based software for control of spectrometer functions and data analysis
- Application templates
- Simple flow bar wizard to create your own methods
- RPF-SQX FP for qualitative and quantitative analysis
- Matching Library for augmentation of FP
- Automatic spectral overlap deconvolution
- Empirical calibration with overlap and matrix compensation
Spectrometer Data
- Single phase AC 100/220V, 15/7A (50/60 Hz)
- Dimensions: 60(W) X 60(D) X 40(H) cm (23.6 X 23.6 X 15.7 inch)
- Weight: 80 kg (176 lbs)
Options
- Fifth secondary target for optimum excitation of Na and Mg
- 9-position automatic sample changer with sample spinner
- Vacuum system
- Uninterruptible power supply (UPS)
- 10-position automatic sample changer
|
|