Home page » products » Agricultural  »  Liquid Chromatography HPLC and UHPLC » Flexar UV/Vis LC and UHPLC detectors
From rugged reliability to outstanding spectral integrity, Flexar UV/Vis detectors offer fast, sensitive performance for a wide range of applications...
Features

From rugged reliability to outstanding spectral integrity, Flexar UV/Vis detectors offer fast, sensitive performance for a wide range of applications. Get every detail about your samples and gain deeper insights in your results.

-       Wide range of flow cells from conventional 10 mm down to high-efficiency micro flow cell for optimum UHPLC peak resolution

-       Captures even the fastest UHPLC peaks with 100 pts/sec detection on FX model

-       Dual-beam optical design with choice of deuterium or tungsten source lamps

 

Specification

Specification

Flexar UV/Vis LC Detector

Flexar FX-UV/Vis UHPLC Detector

CDS control environment

Chromera® and TotalChrom®

Chromera® and TotalChrom®

Wavelength range

190-700 nm

190-700 nm

Bandwidth

5 nm

5 nm

Wavelength accuracy

±1 nm

±1 nm

Optics

Dual beam

Dual beam

Maximum sampling rate

50 pt/sec

100 pt/sec

Sensitivity range

0.0005 to 3.000 AUFS in 0.0001 increments from 0.0005 to 0.1, and 0.01 increments above 0.1 AUFS

0.0005 to 3.000 AUFS in 0.0001 increments from 0.0005 to 0.1, and 0.01 increments above 0.1 AUFS

Noise

<7.5 x 10-6 AU, 210-280 nm, 2 sec response time, standard test cell

<7.5 x 10-6 AU, 210-280 nm, 2 sec response time, standard test cell

Drift

<1 x 10-4 AU/hr, after warmup

<1 x 10-4 AU/hr, after warmup

Light sources

Deuterium (190-360 nm) or Tungsten (360-700 nm), pre-focused, no adjustment required on replacement

Deuterium (190-360 nm) or Tungsten (360-700 nm), pre-focused, no adjustment required on replacement

Flow cell

Flow cell sold separately; compatible with PerkinElmer flow cells: 6 mm x 2.4 μL, 8 mm x 12 μL, 10 mm x 15 μL and 2 mm x 1.7 μL (prep)

2.4 μL, 6 mm pathlength standard; compatible with PerkinElmer flow cells: 6 mm x 2.4 μL, 8 mm x 12 μL, 10 mm x 15 μL and 2 mm x 1.7 μL (prep)

Materials contacting solvent

Kel-F, PEEK, quartz

Kel-F, PEEK, quartz

Leak detection

Heated thermister sensor in glass envelope, located in flowcell drip pan

Heated thermister sensor in glass envelope, located in flowcell drip pan

Dimensions (HxWxD)

16 x 34 x 53 cm

16 x 34 x 53 cm

Weight

11.4 kg

11.4 kg

Print