Atomx XYZ Specifications
Automation
Sample Capacity: 84-positions for 40 mL VOA vials.
Vial Size: Nominal 40 mL capacity, single hole cap with PTFE-faced silicone septum, per EPA specifications; 3 3/4“ high without cap and septum; 1 1/16” OD; 24 mm ID cap for water sampling.
Sample Handling
Liquid Handling: Sample syringe (25 mL) dispenses variable volumes of water from 1 - 25 mL in 0.1 mL increments. 1/16” OD PEEK™ tubing for liquid transfer.
Sample Gas Pathway: Glass, PEEK™ Inertium®, SilcoTek® and PTFE for sample handling.
Cleaning: The entire liquid sample pathway can be rinsed using a combination of the methanol rinse and the high-temperature DI water rinse cleaning techniques. User defined rinse volume and number of rinses for the needle and glassware.
Gas Handling
Electronic MFC: System is capable of controlling flow rates between 5-500 mL/min variable between each mode of operation (patented).
Pressure Monitoring: Ability to record purge and bake pressure for each sample.
Gas Supply: Ultra-high purity (99.999%) Helium or Nitrogen; Incoming Pressure: 65 - 100 psig, (100 psig max)
Standard Injection
Standard Injection: Three standard injection systems utilizing 2-way dosing valves mounted on an internal valve manifold.
Capacity: 1, 2, 5, 10, and 20 μL increments.
Consumption: 1 μL per 1 μL injection
Standard Vessels: Three 15 mL standard vessels, UV-protected and sealed under pressure for standard concentration integrity.
Liquid Samples
(includes drinking water and wastewater; Liquid samples containing up to 15 mm of sediment when measured from the bottom of an upright 40 mL vial)
Sample Glassware: The system is capable of operation with 5-25 mL frit or fritless sparge vessels. Ships standard with 5 mL fritted glassware.
Sample Dilutions: Programmable automatic aqueous sample dilutions of 1:100, 1:50, 1:25, 1:10, 1:5, 1:2.
Blanks: Automatic blanks can be pulled from the water reservoir and spiked with standard allowing all autosampler positions to be used for samples.
Cycle Time: Total Purge & Trap cycle time of 18 minutes, or less depending on the method.
Low-Level Solid Samples
(includes all types of natural soils and sediments. Sampled: Direct purge in the vial per USEPA 5035 low-level soil methodology)
Sample Needle: A patented 3-stage needle allows for water and standards to be directly added to the vial where the solid sample will be purged.
Vial Heater: Variable heat control from 35 °C to 80 °C.
Mixing: The solid sample can be mixed via a stir bar using three variable speeds.
High-Level Solid Samples
(includes all types of soils and sediments. Sampled: Automated Methanol Extraction and subsequent dilution per USEPA 5035 high-level soil methodology)
Extraction: System is capable of fully automating the methanol extraction of high-concentration soil samples.
Matrix Spike: The system is configured to allow a surrogate spike to be added directly to the solid sample when the methanol is added for the extraction.
Extraction Dilutions: Programmable automatic dilutions of methanolic extract of 1:100 or 1:50 for 5 mL sample volumes.
System Control
Instrument Control: Atomx XYZ TekLink™ software in a Windows® 7 or greater environment. Via USB.
Method Scheduling: All method types can be run from any position in the sample sequence. Up to three standards can be added to any user-specified position.
Multiple runs can be made from the same vial (not recommended).
System History: The system records a complete history of all sample, schedule and method information.
Service
Electronic Leak Check: Ability to leak check the entire sample pathway of the system via the automated System Leak Check. Built-in diagnostics that, once a leak has been identified, the system will check independent sub-systems for leaks.
Benchmark Test: The system has a mode that will allow for full electromechanical testing including; valving, heaters, vial handling systems, liquid delivery system, inputs and outputs.
Diagnostics: The system offers independent control of all valves, vial handling mechanisms and syringe drive for troubleshooting.
General Specifications
Dimensions &Weight: Dimensions: 70.1 cm (27.6” W) X 58.42 cm (23” D) X 49.5 cm (19.5” H) Unit weight: 43.1 kg (95 lbs)
Power Requirements: 100-120VAC (±10%), 50/60 Hz, 10.0A, 1150W; 220-240VAC (±10%), 50/60 Hz, 5.0A, 1150W
Environmental Specs: Operating Temperature: 10°-30 °C; Storage Temperature: -20°-60 °C; Relative Humidity: 10%-90%.
Corrosion Resistance: The front cover and sample tray are corrosion resistant to waters with a pH range of 1-10. |