Sensitivity
Test Amount Signal to Noise Performance
EI Full Scan 1 pg of octafluoronaphthalene 650:1
Hardware
Ion Source PerkinElmer SMARTsourceTM—Simplified Maintenance and Removal Technology (EI and CI compatible)
Filament MarathonTM—designed for long life
Mass Range 1.0 – 1,200 u (amu)
Detector ClarifiTM detector—long-life high sensitivity electron multiplier detector
Analyzer Quadrupole with pre-filter
Mass Stability ± 0.1 m/z over 48 hours
EI Voltage 10 – 100 eV
Vacuum Pumps
S model: 75 L/sec (nitrogen) air-cooled turbomolecular pump (59 L/sec helium pumping capacity)
Vacuum Gauge Single wide-range gauge
Pump-down Time 255 L/sec turbomolecular pump: <3 minutes for air/water check, <90 minutes for quantitative stability
Temperature Settings SMARTsource: Settable 50 ˚C – 350 ˚C, GC transfer line: Settable 50 ˚C – 350 ˚C
Carrier Gas Flow Rate 255 L/sec turbomolecular pumping systems: 5 ml/min, 75 L/sec turbomolecular pumping systems: 1.5 ml/min
Mass Calibrant PFTBA (FC-43) gas, triazine (high mass applications), user selectable
MS Isolation SwaferTM Micro-Channel Flow Technology
Field Upgrades Positive/negative chemical ionization (255 L/sec), standard to large turbomolecular pump (75 L/sec to 255 L/sec), water cooling (all models)
Performance
MS Data Collection Full Scan, Selected Ion Monitoring (SIM), Simultaneous Full-Scan Selected Ion Monitoring (SIFI)
Scan Rate Fully variable up to 12,500 amu/sec
Acquisition Rate 100 points/sec (SIM)
Functions/Run 32 functions or 32 ions per function
Linear Dynamic Range 106 dependant on acquisition rate
Data system
Software TurboMassTM 6.0 designed for Windows 7 professional operating system
Methods Electronically transferable between all TurboMass and Clarus GC/MS systems
GC Acquisition Full control and data processing of a single GC/MS with up to 2 GC detectors
MS Tuning UltraTuneTM: Standard Tune environmental tuning optimized for EPA methods; Custom Tune user definable optimization
Reporting Included in all models with specialized reports for: Environmental and forensic applications.
Optional Software MS Libraries: NIST Mass Spectral Library, including AMDIS Deconvolution, Wiley Mass Spectral Library Maurer/Pfleger/Weber Library of Drugs, Pollutants, Pesticides, and Metabolites
Advanced Processing: Ion Signature Deconvolution software
Physical
Power 120 VAC ±10% @ 50/60 Hz ±1% 1,000 VA, 230 VAC ±10% @ 50/60 Hz ±1% 1,000 VA
Temperature 10 ˚C – 30 ˚C (10 ˚C – 35 ˚C with water cooling option)
Relative Humidity 20 – 80 % non-condensing
Weight: 46.7 kg (103 lbs), Forepump: 25.9 kg (57 lbs)
Dimensions (HxWxD)
50 x 32 x 77 cm
|