Analyzer Performance Specifications:
Mass Range 20-3000
Mass Accuracy + 0.1u/24 hr; ± 3ºC
Mass Axis Stability + 0.1u/24 hr; ± 3ºC
Max Scan Rate (u/s) 10,000
Resolution (FWHM) 0.6u ± 0.1 (20-3,000u)
Polarity Switching Standard
Quant. Dyn Rng 10 e5
Ion Sources:
ESI 0.004 - 1.4 mL/min (no split)
APCI 0.05 - 1.5 mL/min (no split)
Dual ESI 0.004 - 1.4 mL/min (no split)
Sensitivity Specifications:
ESI+SIM S/N ≥120:1 RMS 1 pg/ ul reserpine @ 400 ul/ min MeOH/ H2O
ESI-SIM S/N ≥100:1 RMS 2 pg/ ul p-nitrophenol @ 400 ul/ min MeOH/ H2O
APCI+SIM S/N ≥100:1 RMS 1 pg/ul reserpine @ 1.0 ml/min MeOH/ H2O
APCI-SIM S/N ≥100:1 RMS 20 pg/ ul p-nitrophenol @ 1.0 ml/min MeOH/ H2O
Operation, Software and Misc:
Software Chromera®rev. 3 or higher
Operating System XP SP3
Syringe Pump Built -in
Built in Injector/Diverter Optional
Turbo Pump 30/200/200 L/sec
Roughing Pump Edwards® E2M28
Operating 10 ºC to 35 ºC
H, (WxD)= Footprint (cm2):
Without Pump 71H,(34x74) =2.5k
Power Required 230 V
Gas Requirements N2 @100 psi >97% pure, hydrocarbon free; Required ~20-30 L/min |