Home page » products » Liquid Chromatography HPLC and UHPLC » Flexar pumps
More samples done faster. Year after year. That’s what you get when you combine high throughput power with the industry’s most reliable, rugged family of pumps...
Features

More samples done faster. Year after year. That’s what you get when you combine high throughput power with the industry’s most reliable, rugged family of pumps. We have the performance you want, whatever your analytical flow rate needs may be.

-       10x throughput improvement with Flexar FX-10 and FX-15 UHPLC pumps

-       10-15x less solvent consumption for minimal waste disposal, reduced operating costs and greener productivity

-       Proven reliability minimizes cost of ownership

-       Isocratic to binary to quaternary upgrades right in your lab

 

 

Specification

Flexar FX 15

Ultra high performance means the highest resolution, highest sensitivity and fastest analysis – in other words, a superlative LC system. Flexar FX-15 is the ultimate in UHPLC with maximum throughput and minimum stress for you. Getting more sample information in less time lets you be more productive. Just what you want.

-       Dual-reciprocating piston 18,000 psi pump provides as much as 10x or more improvement in productivity

-       Integrated piston wash function keeps pump pistons clean, even with buffers

-       Synchronized injection with piston position for ultimate precision. High-speed FX UV/Vis (100 pt/sec) with high-efficiency 2.4 μL flow cell

-       View pump pressure profiles real-time with Chromera CDS

Flexar FX 10

When the demands in your lab go up, but your budget doesn’t, you need excellent performance at the best possible price. Flexar FX-10’s 10,000 psi capability lets you increase your throughput while keeping your bottom line steady.

-       Minimal disruption in your method development phase and analytical approach so you can easily port your existing methods

-       Micro-binary 10,000 psi pump package

-       At least 5 - 10x throughput improvements

-       High-productivity FX UHPLC autosampler

-       Peltier Column oven for retention time reproducibility and viscosity reduction

-       High-speed FX UV/Vis (100 pt/sec) with high-efficiency 2.4 μL flow cell

 

Flexar HPLC pumps 

Specification

Flexar Isocratic

LC Pump

Flexar Binary

LC Pump

Flexar Quaternary

LC Pump

CDS control environment

Chromera® and TotalChrom®

Chromera® and TotalChrom®

Chromera® and TotalChrom®

Flow-rate range

0.01 to 10.0 mL/min

0.01 to 10.0 mL/min

0.01 to 10.0 mL/min

Flow-rate increments

0.01 from 0 to 0.99 mL/min

0.1 from 1.0 to 10 mL/min

0.01 from 0 to 0.99 mL/min

0.1 from 1.0 to 10 mL/min

0.01 from 0 to 0.99 mL/min

0.1 from 1.0 to 10 mL/min

Flow precision

0.3% RSD (typical 0.1%) at

1 mL/min water @ 1000 psi

0.3% RSD (typical 0.1%) at

1 mL/min water @ 1000 psi

0.3% RSD (typical 0.1%) at

1 mL/min water @ 1000 psi

Flow accuracy

±1% of setting at 1 mL/min

and 1000 psi with water

±1% of setting at 1 mL/min

and 1000 psi with water

±1% of setting at 1 mL/min

and 1000 psi with water

Retention-time reproducibility

<0.3% RSD

(typically <0.1%)

<0.3% RSD

(typically <0.1%)

<0.3% RSD

(typically <0.1%)

Pressure range

0 to 6100 psi for the entire flow-rate range

0 to 6100 psi for the entire flow-rate range

0 to 6100 psi for the entire flow-rate range

Compositional range

N/A

0 to 100%, solvent A to B

0 to 100%, solvent A to D

Compositional accuracy

Typically 0.5% from 3 to 97% up to 5 mL/min

Typically 0.5% from 3 to 97% up to 5 mL/min

Typically 0.5% from 3 to 97% up to 5 mL/min

Compositional precision

Typically <0.2% variation

Typically <0.2% variation

Typically <0.2% variation

Solvent blending increments

N/A

Settable to 0.1% increments

Settable to 0.1% increments

Solvent proportioning valves

N/A

2

4

Gradient forms

N/A

Linear, exponential 1 to 9.9 (positive and negative) or step (189 profile selections)

Linear, exponential 1 to 9.9 (positive and negative) or step (189 profile selections)

Linearity

Within 1% from 10 to 90%

Within 1% from 10 to 90%

Within 1% from 10 to 90%

Program steps

Step 0 (Equil) plus up to 19 flow program steps

Step 0 (Equil) plus up to 19 flow program steps

Step 0 (Equil) plus up to 19 flow program steps

Step time

0 to 999 minutes

0 to 999 minutes

0 to 999 minutes

Step-time increments

0.1 min from 0 to 9.9 min;

1.0 min from 10 to 999 min

0.1 min from 0 to 9.9 min;

1.0 min from 10 to 999 min

0.1 min from 0 to 9.9 min;

1.0 min from 10 to 999 min

Events / method

Max of 9

Max of 9

Max of 9

Dimensions (HxWxD)

16 x 34 x 53 cm

16 x 34 x 53 cm

16 x 34 x 53 cm

Weight

27 kg

27 kg

27 kg

Service Manager software

Pump setup, calibration and troubleshooting using dedicated user interface

Pump setup, calibration and troubleshooting using dedicated user interface

Pump setup, calibration and troubleshooting using dedicated user interface

 

Print