Home page » products » O2 and CO2 Meter in Beverage » Portable CO₂ Meter: CarboQC
CarboQC is the precise CO2 meter which allows reliable QC on finished packages as well as highly accurate measurements in the laboratory for product development. When combined with Anton Paar’s PFD filling device, sample is taken directly from the package into the measuring chamber of CarboQC - with no loss of dissolved carbon dioxide during sample transfer. CarboQC for the laboratory provides reliable CO2 results, even out of very small packages.
Features

        -          CO2 measurement based on the patented Multiple Volume Expansion method guarantees repeatability standard deviations                 of 0.005 vol.

-          No influence of other dissolved gases such as air or nitrogen

-          Low sample volume of around 100 mL provides reliable CO2 results, even out of very small packages

-          CO2 results after only 55 seconds

-          Intuitive user interface with clearly arranged navigation of the menu

-          Clear sample identification for full traceability of 500 measurement results

-          FillingCheck™: automatically detects filling errors and generates a warning message

-          System Check: guides the operator through recommended periodical checks

   -          Auto-valves: Automatically opens and closes inlet/outlet valves by only one Start or Rinsing press

   -          Rinsing Time: Use the rinsing time setting to adjust the transported sample volume for measuring, it is set between 10 and    300 seconds. After that time, inlet/outlet valves will be closed automatically, no need any press of operator

   -          Sample IDs: You can assign a sample ID to each sample. This function helps to uniquely assign measured data to                  samples (sample identification) and the measuring point where the sample was taken, etc. You can save up to 19 different    sample IDs for full traceability of your results. Sample IDs can have up to 10 characters.

   -          Methods: The instrument is equipped with eight predefined methods and 21 additional custom methods.

Fast and wireless

-           Bluetooth interface for wireless printing

-           USB interface for data transfer to a PC

-           RFID interface for changing measurement settings quickly by reading RFID tags

Specification

 Measuring range of CO2:

-          0 g/L to 12 g/L(0 vol. to 6vol.) at 30 °C

-          0g/L to 20 g/L (0 vol. to 10 vol.) < 15 °C

Measuring range of Temperature

-          -3 °C to +40°C (27°F to 104°F),

-          Accuracy: ± 0.2 °C

Measuring range of Pressure: 0 bar to 10bar absolute (0 psi to 145 psi)

Repeatabilitys.d. CO2: g/L (0.005 vol.)

Reproducibilitys.d. CO2: 0.05 g/L (0.025 vol.)

Resolution of CO2: 0.001 g/L

Measuring units: g/L, vol., kg/m3

Sample volume: approx. 100 mL

Measuring timeper sample: approx. 55 seconds

Internalstorage: 500 measured results

Interfaces

-          1 x RS-232 for PC and printer

-          1 x USB for PC

-          1 x RFID, 1 x Bluetooth (optional)

Protectionclass: IP67

Power supply: AC 100 - 240 V, 50/60 Hz,1.5 A; DC 12 V, 3.,0 A

Rechargeablebattery type: Li-Ion 11.1 V, 2.25 Ah(optional: Li-Ion 11.25 V, 2.6 Ah)

Maximumat-line operation: 10 hours