Clarus 480 GC
It is available with manual pneumatic control, singlechannel or dual-channel configurations and optional integral liquid autosampler...
  1. Features
  2. Specification

It is available with manual pneumatic control, singlechannel or dual-channel configurations and optional integral liquid autosampler. Manual pneumatics are available for all injectors, detectors and Two carrier zones. Up to two injectors may be installed and operated simultaneously with independent temperature control. Every injector is available with manual pneumatics. A choice of four detectors, optimized for sensitivity and selectivity, is available for use with the Clarus 480 GC. All built-in detectors include an automated background compensation feature that corrects for column bleed and conform to the highest industry standards for reliability and performance.

 Oven

 ·       Temperature range: 10°C above ambient to 450°C

·       Rate: 0.1 to 45°C/min with increment of  0.1°C

·       Cool-down times: 250°C to 50°C within 4.8 min  

 Autosampler

·       Controlled through the keypad or by a data system such as TotalChrom CDS.

·       Number of sample positions: 108, plus one priority

·       Vial size: 2-mL (0.25mL with insert) crimp-top caps, 2-mL screw-top caps

·       Number of waste and wash vials: 4 waste and 4 wash

·       Waste and wash vial size: 4 mL

·       Syringe size: 0.5μL, 5.0μL or 50.0μL

Injectors  

·       Split/splitless capillary injector

·       Packed column injector

Detectors

·       Flame Ionization Detector (FID)

·       Electron Capture Detector (ECD)

·       Thermal Conductivity Detector (TCD)

·       Nitrogen Phosphorus Detector (NPD)